Detail Syringe Pump Merk JYM Type JZB-1800, Harga Murah
Features
• Compatible with syringes of most of the standard brands.
• Automatic Calibration for syringes.
• Three working modes: Rate control mode, Time control mode and Dose weight control mode ( ug/ kg/ min or mg/ kg/ h) .
• Built in rechargeable battery ( Lithium-ion polymer battery) .
• Alarm functions: occlusion, empty, near empty, low battery, infusion over, syringe loose, etc
• Three levels of the occlusion limited pressure: high, middle and low.
• Three levels of alarm sound: high, middle and low.
• The Syringe pump goes into KVO mode automatically as infusion is completed.
• Purge / Bolus function.
Technical Specification
1. Flow range ( Rate mode) :
50 ml syringe : 0.1 – 1800.0 ml / hr ( 0.1 ml / hr step)
20 ml syringe : 0.1 – 800.0 ml / hr ( 0.1 ml / hr step)
10 ml syringe : 0.1 – 400.0 ml / hr ( 0.1 ml / hr step)
2. Accuracy: + / - 2% ( After correct calibration)
3. Time setting mode range:
Volume: 0.1 – 199.9ml, 0.1ml step
Time: 1 – 1999min, 1min step
4. Weight setting mode range:
Sol. volume: 0.1 – 199.9ml, 0.1ml step
Drug mass: 0.1 – 999.9ml, 0.1 ml step
Dose: 0.01 – 99.99ug/ kg/ min, 0.01ug/ kg/ min step
0.01 – 99.99mg/ kg/ h, 0.01mg/ kg/ h step
Weight: 0.1 – 99.9kg, 0.1kg step
100 – 300kg, 1kg step
5. Purge rate:
1800ml/ h ( 50ml) , 800ml/ h ( 20ml) , 400ml/ h ( 10ml)
6. Volume limits range: 0 – 1999.9ml, 0.1ml step
7. Bolus volume: 1.0 – 100ml, 0.1ml step
8. Bolus Rate:
10ml syringes: 1.0 – 400.0ml/ hr, 0.1ml/ hr step
20ml syringes: 1.0 – 800.0ml/ hr, 0.1ml/ hr step
50ml syringes: 1.0 – 1800.0 ml/ hr, 0.1ml/ hr step
9. Spec of the fuse: F1AL / 250V, two ( inside)
10. Power supply: AC100~ 240V, 50/ 60Hz, Internal Lithium-ion polymer battery: DC 11.1V, e1600mAh. The battery backup is approx. 8 hours ( when using a fully charged battery and running at 5 ml/ h) .
11. Power consumption: d25VA
12. Class: I
13. Type: CF
14. Size: 280mm x 160mm x 130mm
15. Net Weight: 3.5 Kg. ( Approx)
16. Operating condition:
Ambient temperature: 10-40degree C
Relative humidity: 20-90%
17. Storage condition:
Ambient temperature: -10~ 55degree C
Relative humidity: 10~ 90% .
18. Possibility of creation of a database about used solutions ( drugs)